小鼠激活素A(ACV-A)酶聯免疫吸附檢測試劑盒
一鍵復制產品信息
ELK5382
規格: 價格:
48T ¥1820.00
96T ¥2600.00
96T*5 ¥11050.00

Overview 文獻(1)

Product name: Mouse ACV-A(Activin A) ELISA Kit
Reactivity: Mouse
Alternative Names: ACV-A; Activin Beta A Beta A Homodimer
Assay Type: Sandwich
Sensitivity: 5 pg/mL
Standard: 600 pg/mL
Detection Range: 9.38-600 pg/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 3.5h
Research Area: Endocrinology;Reproductive science;Hormone metabolism;
Uniprot ID: Q04998
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse ACVA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse ACVA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse ACVA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse ACVA in the samples is then determined by comparing the OD of the samples to the standard curve.

標準曲線

Concentration (pg/mL) OD Corrected OD
600.00 2.402 2.309
300.00 1.517 1.424
150.00 1.189 1.096
75.00 0.775 0.682
37.50 0.473 0.380
18.75 0.313 0.220
9.38 0.247 0.154
0.00 0.093 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant ACVA and the recovery rates were calculated by comparing the measured value to the expected amount of ACVA in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 82-95% 88%
Heparin plasma(n=5) 95-107% 101%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ACVA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 79-85% 87-96% 82-95% 89-103%
EDTA plasma(n=5) 86-93% 91-105% 87-99% 99-105%
Heparin plasma(n=5) 80-101% 87-96% 96-105% 95-103%